Show simple item record

dc.contributor.authorNadeera Hemamali, GM
dc.contributor.authorSmith, DR
dc.contributor.authorHobson, PR
dc.date.accessioned2024-05-14T15:10:11Z
dc.date.available2024-05-14T15:10:11Z
dc.date.issued2022-01-01
dc.identifier.isbn9781665488723
dc.identifier.urihttps://qmro.qmul.ac.uk/xmlui/handle/123456789/96870
dc.description.abstractIn this paper, a 5×5×5 mm3 YSO:Ce scintillator detector was modelled with FLUKA code, and calculations for the determination of energy resolution are reported using energies of 59.5, 81, 356, and 661.7 keV. We measured the energy resolution of the scintillator detector in the photon energy range of 59.5 to 661.7 keV using radioactive Am-241, Ba-133, and Cs-137 sources and compared the results with the computed energy resolution obtained using FLUKA, the comparison between simulated and measured resolution showing good agreement. This study shows that the FLUKA code can be used to determine the expected energy resolution successfully, saving both financial cost and time in the development of new scintillators and radiation sensors.en_US
dc.publisherIEEEen_US
dc.titleDetermination of Energy Resolution for YSO:Ce Detector Modelled with FLUKA Codeen_US
dc.typeConference Proceedingen_US
dc.rights.holder© 2022 IEEE.
dc.identifier.doi10.1109/NSS/MIC44845.2022.10398959
pubs.notesNot knownen_US
pubs.publication-statusPublisheden_US
rioxxterms.funderDefault funderen_US
rioxxterms.identifier.projectDefault projecten_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record